Clinical careAnesthesia was administered by senior clinicians specializing in neonatal anesthesia.
Infants who arrived in the operating room on a ventilator had preoperatively received morphine sedation at the Neonatal Intensive Care Unit.
Baseline rSO₂ was defined as the median rSO₂ value recorded during the first minute of monitoring before anesthesia induction.
A baseline rSO₂ value was available in 73 of 95 patients, in whom relative changes from baseline were used for event classification.
The timing of each physiological parameter nadir was compared with the rSO₂ nadir to assess its temporal relationship with cerebral desaturation.
Study design and patient population
This prospective observational study was conducted at the Pediatric Center of Semmelweis University, Budapest. The study was approved by the Ethics Committee of the National Medical Research Council (ETT-TUKEB 11790-2/2016/EKU and 7469-12/2021/EUIG). Patients were eligible if they were born between March 2021 and July 2023 and were admitted to our level III neonatal intensive care unit (NICU) with an in-house neonatal surgical center. All patients were outborn. Inclusion criteria were: (1) non-cardiac congenital malformations requiring surgical intervention or complications of prematurity necessitating surgical treatment; (2) availability of multimodal non-invasive intraoperative data recording. We excluded patients with the following conditions: (1) body weight >6 kg at surgery, (2) severe intraventricular hemorrhage or hydrocephalus requiring shunt implantation, (3) technical problems with data recording, and (4) lack of written informed consent from the parent or caregiver. A flow chart of the enrollment procedure and exclusions is provided in Fig. 1. No gestational-age, postmenstrual-age, or postnatal-age restriction was applied. A prespecified body-weight criterion of ≤6 kg at the time of surgery was used to establish a pragmatic cohort of small infants undergoing procedures within our neonatal surgical and anesthetic program. This approach enabled the inclusion of former preterm infants whose definitive or corrective procedures, including stoma closure following necrotizing enterocolitis, were performed beyond the conventional neonatal period.
Fig. 1 Full size image Flow chart of the enrollment procedure.
Clinical care
Anesthesia was administered by senior clinicians specializing in neonatal anesthesia. Induction was performed with inhaled sevoflurane or 1% propofol administered intravenously. Infants who arrived in the operating room on a ventilator had preoperatively received morphine sedation at the Neonatal Intensive Care Unit. During induction, fentanyl and muscle relaxants were administered as required, likewise, regional anesthesia. All patients were mechanically ventilated throughout the procedures in accordance with our institutional protocol. Airway management (laryngeal mask airway or endotracheal tube) and ventilation mode (pressure support or pressure-controlled ventilation) were left to the anesthesiologist’s discretion. At the end of the procedure, extubation was at the discretion of the anesthesiologist. If the attempt proved unsuccessful or was not intended at the end of surgery, the patient remained on mechanical ventilation.
Cardiovascular management aimed at maintaining the mean arterial blood pressure at or above the infant’s gestational age in weeks. Hypotension was treated with fluid boluses or vasopressor agents, determined by the clinician.
During the surgical procedure, standard anesthetic monitoring was used, consisting of a pulse oximeter, electrocardiogram, non-invasive blood pressure monitor, thermometer, and side-stream capnograph, and was complemented by a NIRS monitor. Clinicians could view cerebral oxygenation values, but clinical decisions were not based on them.
Equipment and data retrieval
A multimodal monitoring setup was assembled for research purposes to complement routine clinical care in the operating room. The anesthesia workstation (Perseus A500, Dräger, Lübeck, Germany) and the patient monitor (Infinity DeltaXL, Dräger, Lübeck, Germany) were connected to a laptop using a recording software developed by the Research and Development Department of Dräger (Lübeck, Germany) to download synchronized time-series data of routinely monitored physiological parameters. This tool is intended exclusively for scientific purposes, and it is not commercially available.
For the present study, parameters included peripheral oxygen saturation (SpO 2 ), heart rate (HR), non-invasive systolic, diastolic, and mean arterial blood pressure (MAP), and end-tidal carbon dioxide (etCO 2 , obtained with a side-stream capnograph), all collected from the anesthesia workstation or the Infinity patient monitor. Cerebral oxygenation was monitored using the Invos 5100 C NIRS (Medtronic, Dublin, Ireland) device, with dedicated neonatal sensors provided by the manufacturer. Raw rSO 2 were sampled at 0.2 Hz frequency and exported at 0.033 Hz (one sample every 30 seconds). Continuous physiological variables, including rSO₂, HR, SpO₂, and etCO₂, were summarized as 1-minute median values. MAP was routinely measured at 5-minute intervals during clinically stable periods. For synchronized temporal analysis, each MAP measurement was carried forward until the subsequent measurement. If the blood pressure measurement interval was shortened because of clinical indications, all additional MAP measurements were incorporated into the analysis. Acquisition frequencies for all monitored variables are summarized in Supplemental Table S1. Time synchronization was performed at the beginning of each recording session. The anesthesia workstation and patient monitor used synchronized internal clocks, which were routinely verified, particularly after daylight saving time changes. The internal clock of the INVOS™ 5100 C monitor was manually aligned with the study computer before each recording because no electronic synchronization interface was available at the time of the study. This procedure provided synchronization with a resolution of approximately one minute. Given that all physiological variables were summarized as 1-minute median values prior to analysis, this temporal resolution was considered appropriate for the analyses performed.
Data processing and statistical analysis
Data processing was performed using Python (version 3.13, https://www.python.org). For the present analysis, 1-minute median values were generated for each parameter and aligned by timestamp.
Descriptive statistics were reported as mean ± standard deviation or median with interquartile ranges [IQR] for continuous variables and as numbers with percentage (%) for categorical variables. Fisher’s exact test and the Wilcoxon rank sum test were used to compare categorical and continuous variables, respectively.
Whenever feasible, NIRS recording was initiated before the induction of anesthesia in non-sedated infants. Baseline rSO₂ was defined as the median rSO₂ value recorded during the first minute of monitoring before anesthesia induction. However, a pre-induction baseline could not be obtained in all patients because NIRS sensors were applied after induction of anesthesia or because the infant arrived in the operating room already sedated and mechanically ventilated. A baseline rSO₂ value was available in 73 of 95 patients, in whom relative changes from baseline were used for event classification. In the remaining 22 patients, predefined absolute rSO₂ thresholds were used. Events with rSO 2 decreases were categorized according to previously published criteria.10,17 Briefly, a mild decrease was defined either as an 11-20% decrease from the baseline or an absolute rSO 2 value of 60-69 (if the baseline value was unavailable). A moderate decrease was considered as a 21-30% decrease from the baseline or an absolute rSO 2 value between 50-59. Events were considered severe if there was a decrease of more than 30% from the baseline or the absolute rSO 2 value was below 50. Event severity (mild, moderate, or severe) was determined at the rSO₂ nadir using the percentage decrease from the individual baseline when available, or the predefined absolute rSO₂ thresholds when baseline values were unavailable.
Event detection was performed on 1-minute median rSO₂ values rather than on the raw NIRS signal, thereby reducing the influence of transient motion artifacts and brief signal losses. In addition, all recordings underwent manual quality review, and obvious non-physiological signal artifacts were excluded before event identification. Potentially unreliable etCO₂ values were manually reviewed by visualizing the corresponding capnogram waveforms. Values were considered valid only when a recognizable expiratory plateau allowed reliable end-tidal CO₂ determination; measurements without an adequate capnogram morphology were excluded.
Ventilation parameters, including minute ventilation, respiratory rate, and tidal volume, were also examined to ensure accurate interpretation of etCO₂ findings.
Missing rSO₂ intervals ≤4 minutes were excluded when both preceding and subsequent values were within the normal range. When values flanking a gap indicated ongoing desaturation, gaps ≤4 minutes were considered part of the same event, whereas desaturation periods separated by gaps >4 minutes were considered separate events. The 4-minute threshold was selected pragmatically to balance the avoidance of event fragmentation with limiting assumptions of continuity across missing data. It was chosen to remain shorter than the 5-minute non-invasive blood-pressure sampling interval.
RSO 2 -time integral
To quantify the cumulative burden of cerebral desaturation, the rSO₂-time integral (area below the individual baseline and above the rSO₂ curve; Fig. 2) was calculated for each mild, moderate, and severe event. This metric was analyzed separately from the event severity classification.
Fig. 2 : The rSO 2 -time integral concept. Full size image A Schematic representation of the rSO 2 -time integral concept. B Representative case depicting rSO 2 changes during esophageal atresia repair of a 12-day-old female infant. C Representative case depicting rSO 2 changes during laser photocoagulation for retinopathy of prematurity in a 2.5-month-old male infant born at 24 weeks of gestation.
Temporal association of rSO 2 and other physiological parameters
Severe rSO₂ events were analyzed to investigate the temporal relationship between cerebral oxygenation and routinely monitored physiological parameters (HR, MAP, SpO₂, and etCO₂), as these extreme events were considered most likely to reveal physiologically relevant temporal associations. The nadir values of each parameter were identified and compared with a reference value obtained 5 minutes before the event, provided that the reference value was within the normal range. The Wilcoxon signed-rank test was used to compare nadir and reference values. Percentage and absolute changes from the reference value were calculated. The timing of each physiological parameter nadir was compared with the rSO₂ nadir to assess its temporal relationship with cerebral desaturation. In addition, cerebral fractional tissue oxygen extraction (cFTOE) was calculated as (SpO₂ − rSO₂)/SpO₂. cFTOE was determined at the reference time point 5 minutes before severe event onset and at the rSO₂ nadir and was analyzed descriptively. Statistical analyses were performed in R (v4.1.2; R Foundation for Statistical Computing, Vienna, Austria).