The axial localization of a fluorophore reported here necessitated a considerable expansion of the MINSTED setup over previous 2D-MINSTED implementations11, involving additional STED lasers and positioning systems. The confocal scanner addresses a field of 64 µm × 64 µm in the lateral plane of the sample with two galvo mirrors12. When at least N on photons were detected in a 2 by 2 pixels neighborhood, a 3D-MINSTED localization was initiated11. Measurements were performed on four different cells (Supplementary Fig. For the retained tracks, the localization precisions \({\sigma }_{{xy}}\left({N}_{{xy}}\right)\) and \({\sigma }_{z}({N}_{z})\) were calculated as described above, based on the rotated data.

Experimental setup

The experimental setup is sketched in Supplementary Fig. S3, and its components are listed in the Supplementary Information. The axial localization of a fluorophore reported here necessitated a considerable expansion of the MINSTED setup over previous 2D-MINSTED implementations11, involving additional STED lasers and positioning systems. In our setup, the lateral MINSTED localization is realized by overlapping the beam of a ps-pulsed excitation laser emitting at a wavelength of 532 nm with the beam of a self-built ns-pulsed fiber-amplified STED laser emitting at 636 nm (636 1 ); by deflecting those beams in the pupil planes with two electro-optical deflectors (EOD x,y ); and by circling their foci around the estimated position of the fluorophore at 125 kHz frequency with a maximum radius of 125 nm. The STED beam passes through a vortex-phase mask in the pupil plane to create the STED donut in the focal plane for lateral emission confinement. After passing the EODs, both beams are p-polarized and reflected by a deformable mirror. By modulating the wavefront of the beams23, the deformable mirror is used to scan a field of up to 3 µm × 3 µm × 3 µm extent and shift the center position in 3D during the localization process. An achromatic quarter-wave plate (λ/4) establishes circular polarization for all beams before being focused into the sample by a silicon oil immersion objective lens of 1.3 numerical aperture (NA). The fluorescence signal is descanned by the deformable mirror, reflected at the polarization-sensitive dichroic mirrors (DM 2,3 ), and is detected with a single-photon avalanche diode (APD 1 ) at 550–610 nm wavelength.

For confinement of fluorescence in the axial direction, two additional ns-pulsed STED beams at 636 nm wavelength with mutually orthogonal polarization (636 2,3 ) pass through a segmented achromatic half-wave plate (λ/2 s ), ensuring p-polarization on the SLM. The SLM controls the position and shape of the foci of these additional beams. Phase profiles ensuring a top-hat phase modification and an axial defocus create 3D donuts above and below the focal plane. The SLM allows for aberration correction in the sample24. The calibration of the defocus phase profile is described in the Supplementary Information. The beams for 3D donut formation are joined with a Wollaston prism (W 1 ). The common beam is transmitted through an achromatic quarter-wave plate, a resonant electro-optical modulator (Res-EOM), driven at a frequency of 20 MHz, and a polarizing beam splitter (PBS) to clean up the polarization. The Res-EOM periodically rotates s-polarized light to p-polarization, thereby allowing to combine alternating pulses of the orthogonally polarized beams into one p-polarized beam. Afterwards, the beam transmits through DM 3 , is rotated to s-polarization by an achromatic half-wave plate (λ/2), and is reflected by DM 2 onto the deformable mirror.

The power of each pulsed laser is controlled by acousto-optical modulators (AOMs). The timing of the pulses for the lasers and the analog sinusoidal wave for the Res-EOM are defined by a static pattern, provided by an arbitrary wave generator.

The setup is based on confocal beam scanning, which can be coupled into the system with a flip mirror. The confocal scanner addresses a field of 64 µm × 64 µm in the lateral plane of the sample with two galvo mirrors12. Continuous-wave lasers provide excitation light at a wavelength of 488 nm and 560 nm with a fixed power of around 15 µW in the aperture of the objective lens. The detection part of the setup splits the emitted fluorescence into two color channels of 495–520 nm and 580–610 nm wavelength ranges.

For calibrating the defocus phase profile (Supplementary Fig. S12), a pellicle is manually placed into the beam in order to deflect a part of outcoming light onto a photomultiplier tube (PMT). Due to the different polarization of the STED beams, the PBS can be rotated to select the corresponding polarization.

To select the region of interest (ROI), the sample is mounted on a piezo stage for fine movements. The piezo stage is mounted on three linear stages for coarse positioning. The sample can be actively stabilized both laterally and axially. The axial feedback signal is obtained by detecting the reflected spot of a power-stabilized super-luminescent light-emitting diode (SLED) at 1000 nm wavelength from the coverslip sample onto the camera CAM 2 . The lateral feedback signal is obtained by tracking fiducial markers in darkfield images on the camera CAM 1 . The fiducial markers are imaged in a field of 70 µm × 70 µm. They are placed off-axis to avoid interactions with the MINSTED illumination of the sample. For imaging the fiducial markers, the reflected SLED illumination beam is blocked by an iris diaphragm. The fundamental optical structure and the software of the focus locks are described in previous work12,25.

The microscope is controlled by two field-programmable gate array (FPGA) devices and a custom graphical user interface implemented in LabVIEW 2021 (National Instruments) and MATLAB R2021b (The MathWorks, Natick, MA, USA).

MINSTED measurements

A fluorescent molecule was searched for by applying a confocal scan in a lateral ROI using the deformable mirror. When at least N on photons were detected in a 2 by 2 pixels neighborhood, a 3D-MINSTED localization was initiated11.

The lateral localizations were executed with a sampling pattern radius \({R}_{{xy}}\) equal to half the FWHM of the lateral E-PSF. Upon each lateral photon detection, the lateral center \(\left(x,y\right)\) of the pattern was moved \({\alpha }_{{xy}}{R}_{{xy}}\) towards the causative sampling position; the STED energy was increased such that the FWHM of the lateral E-PSF was reduced to the radius \({\gamma }_{{xy}}{R}_{{xy}}\) down to a preset minimum. We used \({\alpha }_{{xy}}=15\%\) and \({\gamma }_{{xy}}=97\%\) as established previously11,12,13,23. The axial localizations were performed with a fixed axial sampling at \(\pm {R}_{z,{\small{\mbox{samp}}}}\) equal to 0.75 R z (for imaging) and 0.5 R z (for tracking), where \({R}_{z}\) is half of the minimal FWHM of the axial E-PSF. Upon each axial photon detection, the axial center z of the sampling pattern was moved \({\alpha }_{z}{R}_{z}\) towards the causative sampling position, and the FWHM of the axial E-PSFs was reduced to \({\gamma }_{z}{R}_{z}\) down to a preset minimum. Supplementary Fig. S13 shows the correlation between α z and the simulated localization precision in the axial direction \({\sigma }_{z}\). We typically applied \({\alpha }_{z}=10\%\) and a fraction \({\gamma }_{z}=97\%\), which proved robust at low SBR. The minimal STED pulse energies were \({E}_{{xy}}=0\) nJ and \({E}_{z}=0.01\) nJ and corresponded to minimal radii \({R}_{{xy}}=118\) nm and \({R}_{z}=200\) nm. Maximal pulse energies of \({E}_{{xy}}=0.62\) nJ and \({E}_{z}=0.4\) nJ defined a scan radius of \({R}_{{xy}}\ge 18\) nm in the lateral plane and \({R}_{z}\approx 40\) nm along the axial direction. Both lateral and axial localizations were performed simultaneously by cycling through the three STED beams pulse by pulse. Each excitation and STED pulse pair was placed in a separate interval to allow for an unambiguous attribution of the detected photons (Fig. 1c–f). The 3D-MINSTED localization was terminated if less than \({N}_{{\small{\mbox{off}}}}=16\) photons were detected during an interval time \({t}_{{\small{\mbox{off}}}}=8\) ms.

The DNA origami samples, ordered from the manufacturer GATTAQuant, were prepared with an origami dilution of 1:9 in 10 mM MgCL 2 in PBS on coverslips. Gold beads of 200 nm diameter (EM.GC200/7, BBI Solutions) were placed on and served as fiducial markers on the coverslip for the lateral focus lock. The origamis were imaged with DNA-PAINT using a Cy3B-labeled imager strand concentration of 7.5 nM. After 5 min of incubation time, the unbound origamis were washed out four times with 75 mM MgCl 2 in PBS. Afterwards, the sample was sealed in a solution of 200 μL of oxygen-deprived reducing-oxidizing buffer with a MgCl 2 concentration of 75 mM26. A more detailed description of the sample preparation is described in the literature12. To ensure that the DNA-PAINT strands were bound to the DNA origami grid, a search dwell time of 1.6 ms per 80 nm pixel and a threshold of \({N}_{{\small{\mbox{on}}}}=40\) photons were chosen. The sample was actively stabilized in all directions because the measurement time for a single FOV was around 40 min.

For the tracking experiments, the search was performed with a dwell time of 60 µs per 75 nm pixel. The threshold was set to \({N}_{{\small{\mbox{on}}}}=2\) photons to measure as many potential tracks as possible for a-posteriori filtering during the data analysis. Measurements were performed on four different cells (Supplementary Fig. S8) with different FOVs per cell. Here, gold beads could not be used as fiducial markers due to their incompatibility with cells. Hence, only the axial direction was actively stabilized. For a single track, no significant drift was expected since the maximum measured time for a single track was 2.4 s (Supplementary Video 1).

Kinesin labeling

Cysteine-light truncated (at amino acid position 560) human kinesin-1 with a single solvent-exposed cysteine at amino acid position 28 was expressed in Escherichia coli using the plasmids K560CLM K28C that were purified as described in a previous study8. Kinesin-1 was labeled with Cy3B maleimide (19380, Lumiprobe) overnight at 4 °C. Excess dye was removed using ZEBA spin desalting columns 7 K MWCO (89877, Thermo Fisher Scientific). The concentration and the degree of labeling efficiency were measured using UV-vis spectroscopy and mass spectrometry. Labeled kinesin was supplemented with 10 % (w/v) sucrose and stored at −80 °C until further use.

Cell culture

U-2 OS cells (European Collection of Authenticated Cell Cultures, cat. 92022711, lot 17E015) were cultured in McCoys media (16600082, Thermo Fisher Scientific) supplemented with FBS Superior (S 0615, Bio&SELL), Pen/ Strep, (P0781, Sigma Aldrich), Natirum-Pyruvat (S8636, Sigma Aldrich) and MEM NEAA (M7145-100ML, Sigma Aldrich). The cells were kept at 37 °C, 5 % CO 2 , and 100 % humidity.

Fixed-cell kinesin tracking

U-2 OS cells were seeded at least 1 day prior to experiments on 18 mm round microscopy coverslips. For fixation, the cells were incubated in extraction buffer (PEM80: 80 mM PIPES, 1 mM EGTA, 1 mM MgCl 2 set to pH 7.4 supplemented with 1 mM sucrose and 0.15% Triton X-100) for precisely 1 min. Next, an equal amount of 2 % (w/v) (1 % (w/v) final concentration) paraformaldehyde in PEM80 was added and incubated for 1 min. The solution was removed and the cells were washed 3 times for 1 min with PEM80 supplemented with 1 µM Paclitaxel (10-2095, Focus Biomolecules).

Finally, imaging buffer was applied to cavity slides. The imaging buffer contained ~2.5–5 nM labeled kinesin, 500 µM ATP, 1 mM DTT, 1 mM methyl viologen, 1 mM ascorbic acid, 20 µM Paclitaxel, 0.4 µg/µL Pyranose Oxidase (P4234-250UN, Thermo Fisher Scientific), ~96 U/mL Catalase (Bovine, 100429, MP Biomedical) in PEM80 supplemented with 10 % (w/v) glucose. The coverslip was then mounted with cells facing towards the cavity and sealed using Picodent Twinsil Speed 22 (Picodent Dental-Produktions- und Vertriebs GmbH) and curated at 37 °C for 5 min.

Live-cell kinesin tracking

U-2 OS cells were seeded at least two days prior to the experiments. The following day after seeding, 500 ng plasmid DNA HaloTag-KIF5B(1–560) (#213403, Addgene) in Opti-MEM was mixed with 2 µL lipofectamine 2000 (11668019, Thermo Fisher Scientific) and incubated for 15 min. Afterwards, the mixture was applied dropwise to the cell culture. After 2 h, the media was exchanged with fresh, prewarmed FluoroBrite DMEM (A1896701, Thermo Fisher Scientific), and the cells were left overnight to incubate.

On the experiment day, cells were labeled with 100 nM JF549 HaloTag ligand (HT1020, Promega) for 15 min in fresh, prewarmed FluoroBrite. Then, the cells were washed 2 times for 15 min in FluoroBrite supplemented with 10 µM Paclitaxel. Afterwards, the coverslips were mounted and sealed as described previously.

Data analysis

The data analysis was designed by us and was implemented and executed in MATLAB R2021b. Some functions were inspired by ChatGPT code examples, in particular for the visualization of the data.

DNA origami evaluation

Localization tracks with a photon number \(N > 3000\) were split into tracks spanning a maximum photon number of \({N}_{\max }=3000\). Initial photon detections N c used to reach the minimum FWHM of the respective E-PSFs were discarded. To ensure static single fluorophores at one binding site, the tracks were filtered by a range of the mean detection rate; a minimal number of detections per track; a maximum standard deviation in each dimension; and a maximum ellipticity in the lateral plane. The parameter histograms of the unfiltered and filtered tracks are shown in Supplementary Fig. S4. Supplementary Table 1 lists the corresponding parameter limits.

Due to separate lateral and axial updates in our 3D-MINSTED localization process, the center positions \({\left({x}_{i},{y}_{i},{z}_{i}\right)}_{i\ge {N}_{c}}\) were split into sequences of j updates in the lateral plane \(({x}_{j},{y}_{j})\) and along the axial direction \(({z}_{j})\). The localization precisions were obtained by calculating the standard deviation \({\sigma }_{{xy}}\left({N}_{{xy}}\right)=\sqrt{{\sigma }_{x}\left({N}_{{xy}}\right)\cdot {\sigma }_{y}\left({N}_{{xy}}\right)}\) and \({\sigma }_{z}({N}_{z})\) of the center positions averaged in blocks of \({N}_{{xy}}\) photons for \(({x}_{j},{y}_{j})\) and, respectively, \({N}_{z}\) photons for \(({z}_{j})\). For each valid localization precision estimate, we required at least 5 non-overlapping blocks. To ensure comparable results across the full range of photon numbers, tracks with no valid localization precision at \({N}_{\min }=200\) photons (equivalent of \(5{N}_{\min }=\)1000 photons in both axial and lateral directions, respectively) were discarded. For the subsequent evaluation, the mean values of center positions \(\left(\bar{x},\bar{y},\bar{z}\right)\) were used for clustering and fitting.

To estimate the binding sites, the retained center positions \(\left(\bar{x},\bar{y},\bar{z}\right)\) were grouped in clusters with the function “dbscan” (MATLAB 2021b), maximally covering 8 nm in diameter in 3D. A secondary clustering was then performed with a maximum diameter of 34 nm to detect the 3 × 3 DNA origami grids. GATTAQuant specifies them with a grid constant of 12 nm along one direction and 10.9 nm along the orthogonal direction. Clusters with less than six clustered binding sites and the corresponding tracks were discarded.

For each origami cluster, a 3 × 3 grid representing the expected binding sites was matched to the measured cluster to get the cluster’s orientation in 3D and to estimate its grid constants. Next, the center positions were assigned to the nearest expected binding site, where center positions from the same track were allocated together to a cluster and binding site. Again, clusters with less than six binding sites and the corresponding tracks were discarded, resulting in a total of 87 matched and retained origamis. To overlay the matched clusters, they were recentered and rotated into a common orientation \(\left({x}_{{\small{\mbox{rot}}}},{y}_{{\small{\mbox{rot}}}},{z}_{{\small{\mbox{rot}}}}\right)\), where the larger grid constant was oriented along the x-axis and the smaller one along the y-axis. The retained clusters are shown in Supplementary Fig. S5.

For the retained tracks, the localization precisions \({\sigma }_{{xy}}\left({N}_{{xy}}\right)\) and \({\sigma }_{z}({N}_{z})\) were calculated as described above, based on the rotated data. Modeling those values with the function \({\sigma }_{{\small{\mbox{est}}}}(N)={a/(b+N)}^{c}\) \(0 < a,\,0 < b\) and \(0 < c < 0.5\) estimates the lateral and axial localization precisions for the complete track12.

Tracking evaluation

Tracks with at least 1000 detected photons and a standard deviation of at least 10 nm along the x- or y-axis were considered as tracks potentially stemming from moving kinesin. The initial detections before the E-PSFs reached their smallest values; detections during a further 20 ms interval; and detections during the final 20 ms were ignored to avoid intervals with insufficient localization precision and potential excursions due to background detections. The count rate was estimated as the reciprocal of the moving mean over time differences of consecutive photon detections with 100 binned values. Overly bright tracks with a mean count rate >16,000 cps (counts per second) were removed because several fluorophores might have been tracked simultaneously. Tracks with a smaller count rate are probably either single fluorophore tracks or two fluorophore tracks with a bleaching event. To find a potential bleaching event from two to one fluorophore, an edge detection on the count rate was performed by using the function “ischange” (MATLAB R2021b) with the “Statistic” parameter set to “mean” and the penalty parameter “MinThreshold” =\(1/3\cdot n\cdot {{{{{{\mathrm{var}}}}}}}\left\{{{{{{{{\rm{countrate}}}}}}}}\right\}\), where n denotes the number of count rate values in the track. If more than one bleaching step was found for a track, the track was removed, because it could stem from multiple fluorophores that were not removed by the previous mean count rate filter. Detections within the first 20 ms after the bleaching step were also removed to keep only single fluorophore tracks for further analysis.

A principal component analysis was performed on the center positions of the retained fluorophore tracks, and the coordinates were rotated along the principal axis in the lateral plane with a new coordinate system \(\left({x}_{{{{{{{{\rm{M}}}}}}}}},{y}_{{{{{{{{\rm{M}}}}}}}}},\,{z}_{{{{{{{{\rm{M}}}}}}}}}\right)\), where the x M -axis is the direction of the principal axis. In order to remove static fluorophores, tracks briefer than \({t}_{\min }=\) 100 ms and tracks spanning less than 80 nm along the principal axis (i.e., less than 5 potential kinesin steps) were filtered out. Particularly for the live cell data, the trajectories along the microtubules were not linear, probably due to the flexibility and bending of the microtubules. Therefore, the shape of the microtubule was modeled with a third-degree polynomial along the orthogonal axes y M and z M . The standard deviation of the distance from the microtubule model was calculated to obtain an initial estimate of the single-photon spatial precisions \({\sigma }_{{\small{\mbox{lat}}}_{{\tiny{{\rm{M}}}}},{\small{\mbox{est}}}}\) and \({\sigma }_{{z}_{{\tiny{\mbox{M}}}},{\small{\mbox{est}}}}\) on the curved trajectory. Due to the rotational symmetry of the vortex donut, the single-photon spatial precisions along the principal axis \({\sigma }_{{x}_{{\tiny{{\rm{M}}}}},{\small{{\rm{est}}}}}\) and orthogonal axis \({\sigma }_{{y}_{{\tiny{{\rm{M}}}}},{\small{{\rm{est}}}}}\) were assumed to be equal to \({\sigma }_{{\small{{\rm{lat}}}}_{{\tiny{{\rm{M}}}}},{\small{{\rm{est}}}}}\). Tracks with large single-photon spatial precision \({\sigma }_{i,{\small{{\rm{est}}}}} > {\sigma }_{\max }\) were filtered out. Such tracks may have been disturbed by other fluorophores moving through the sampling volume. The unfiltered and filtered tracks, along with the corresponding filter parameters, are listed in Supplementary Table 2 and shown in Supplementary Figs. S7 and S9.

To identify kinesin steps, the retained tracks were segmented with a custom “pruned-exact-linear-time” (PELT) algorithm27,28 on the center positions \(\left({x}_{{{{{{{{\rm{M}}}}}}}},i},\,{y}_{{{{{{{{\rm{M}}}}}}}},i},\,{z}_{{{{{{{{\rm{M}}}}}}}},i}\right)\), where the centers were normalized with the corresponding single-photon spatial precision in the corresponding axis. The PELT penalty parameter was set to 120 normalized spreads. Segments spanning less than 120 photons were flagged as unreliable due to large position uncertainty. Segments were also flagged as unreliable if the distance between the mean positions in segments was <5 nm, probably caused by (i) just the localization uncertainty, (ii) several kinesin steps in rapid succession, leading to distances >32 nm, or (iii) if the step direction was not aligned with the principal axis. Finally, only tracks with at least two distance values were considered. The results for all segments and for sequential reliable segments only are shown in Supplementary Figs. S14 and S15, and the corresponding filter parameters are given in Supplementary Table 3.

The spatiotemporal resolution and the estimation of the SBR are introduced in reference13 and briefly described in the Supplementary Information.

Statistics and reproducibility

The measurements for the 3 × 3 DNA origami grids were performed for 7 different FOVs, yielding 17,259 raw tracks. After applying the first filter to ensure that only single-fluorophore tracks were kept, 13,511 tracks remained. Of these, 3787 tracks could be sorted into a grid with at least six binding sites, resulting in a total of 599 different binding sites that were investigated across 87 grids for the 7 different FOVs. Figure 2b, c displays grids of one exemplary FOV. Figure 2h shows an overlay of all 87 detected grids with localizations based on 200 binned photons.

For the tracking data shown in Figs. 3 and 4, 7 fixated and 4 living cells with 18 (Supplementary Fig. S16) and 20 (Supplementary Fig. S17) FOVs were measured, resulting in 188 and 337 evaluated tracks with 788 and 1207 steps, respectively.

Reporting summary

Further information on research design is available in the Nature Portfolio Reporting Summary linked to this article.

© All Rights Reserved.